    Precautions for the Use of 3S-J5000 Ozone Online monitor

    Views :
    Update time : 2025-08-08

    Precautions for the Use of 3S-J5000 Ozone Online Mointor


    First, introduce air or oxygen, then turn on the detector. The detector will automatically zero. During the mid-process zero calibration, there must be gas passing through (whether it is air, oxygen or ozone).


    If ozone remains in the system normally, please switch to manual mode for zero calibration


    ■ There must be anti-backflow measures. It is strictly prohibited for water to flow back into the detector!


    The intake pipe of the detector must be made of polytetrafluoroethylene, 316L stainless steel or ozone-resistant materials; otherwise, the pipe will oxidize and fall off, contaminating the detection light cell!


    ■ Preheat the detector for 5 minutes before use!


    ■ The experiment time is short. It is recommended to change to manual zeroing or automatic zeroing, with a time of 30 minutes!


    ■ Ozone water treatment and ozone tail gas detection require the addition of a drying device!


    ■ The intake air volume is adjusted to be around 0.5L/min, and the intake air pressure is ≤0.1MPa


    Turn on the oxygen/air first, then the ozone tester, and then the ozone generator


    ■ First turn off the ozone generator and the ozone tester, then the oxygen. (Reversing the sequence will cause ozone residue in the detector and inaccurate zero calibration for the next use.)


    ■ The intake air must not contain impurity gases such as ammonia and nitrogen oxides; otherwise, it will contaminate the photocell.



    下一篇 :
    上一篇 :
    Related News
    Read More >>
    Introduction to the Light Sour Introduction to the Light Sour
    08 .05.2026
    Introduction to the Light Source Structure of UV Ozone AnalyzersUV absorption is
    Online vs Offline Dissolved Oz Online vs Offline Dissolved Oz
    08 .05.2026
    Online vs Offline Dissolved Ozone Monitoring: Pros and ConsDissolved ozone is wi
    Ozone Analyzer Calibration Int Ozone Analyzer Calibration Int
    07 .31.2026
    Ozone analyzers are calibrated at the factory, but sensors gradually drift duri
    Understanding Ozone Concentrat Understanding Ozone Concentrat
    07 .31.2026
    Understanding Ozone Concentration Measurement Units: ppm, ppb, mg/L, and Ug/m3;I
    網站地圖